Follow these quick-start checklists each time you use the IFL (this assumes you've already successfully set up your system).
Set up camera masks
Remove everything from the capture space, including yourself, the calibration wand, and your tracking objects (anything reflecting IR will be masked out by the cameras and thus will create dead spots)
In Vicon Tracker go to "Calibrate">"Create Camera Masks">"Start"
Watch the camera overview panes gradually mask any IR-reflecting pixels from green to blue. Stop once finished.
Calibrate the space (do this every session once the cameras have reached thermal equilibrium)
In Vicon Tracker go to "Calibrate">"Calibrate Cameras">"Show Advanced" and set the following:
Wand: 5-Marker Wand and L-Frame
Calibration type: Full Calibration on All Cameras
Refinement frames: At least 2000, the more the better!
Tick "Auto-Stop"
Click "Start"
Do the "Vicon wand dance" being sure to maximize the simultaneous visibility of all cameras
Once the calibration is completed check the "Camera Calibration Feedback" and repeat calibration if necessary. It is totally feasible to have less than 1mm World Error from all cameras
"Set Volume Origin" by placing the wand at the desired position and axis orientation and clicking "Start"
You are now ready to commence a tracking task!
At the end of your session be sure to:
Shut down Vicon Tracker as you will block the license for the next user if not
Unplug the orange extension cable
Switch off the surge protector strip
Report any issues to Lee Whitcher
Remove all trash and all of your equipment